Data Foundry

Medical device recalls 2022

Hillrom Hospital Bed Accessory — Class I medical device recall Z-0142-2023

Ongoing recall by Baxter Healthcare Corporation, reported 2022-11-09. Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including,

Recall numberZ-0142-2023
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBaxter Healthcare Corporation, Deerfield, IL, United States
Recall initiated2022-09-30
Classified2022-10-31
Reported by FDA2022-11-09
Quantity814 units
UPC / GTIN / UDI-DI00887761984998
Model numbersP00697905

Product

Hillrom Hospital Bed Accessory, Watch Care Incontinence Management System for Centrella Bed, Product Code/Part Number P00697905

Reason for recall

Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0142-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.