Data Foundry

Medical device recalls 2025

Brand Name: ARVIS Product Name: ARVIS" Shoulder Model/Catalog Number:… — Class II medical device recall Z-0145-2026

Ongoing recall by Kico Knee Innovation Company, reported 2025-10-22, distributed nationwide. Complaint identified issue with AI surgical planning software that may result in implant malalignment and/or d

Recall numberZ-0145-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmKico Knee Innovation Company, Pymble, Australia
Recall initiated2025-09-19
Classified2025-10-14
Reported by FDA2025-10-22
DistributedNationwide (US)
Quantity21 systems
Reason classessoftware
Model numbersIN-27300

Product

Brand Name: ARVIS Product Name: ARVIS" Shoulder Model/Catalog Number: IN-27300, Arvis Surgical Planning Software Software Version: V2025.1.2 Product Description: Intended Use: The ARVIS" system is a computer-controlled navigation system intended to provide intra-operative measurements to the surgeon to aid in selection and positioning of orthopedic implant components. Description: ARVIS Shoulder is a computer-controlled surgical navigation system for shoulder arthroplasty. It aids the surgeon in making intra-operative measurements and locating anatomical structures of the shoulder joint based on the patient s preoperative imaging to assist with selection and positioning of orthopedic implant components. The system consists of software, electronic hardware and surgical instruments. The device s workflow involves CT based preoperative planning followed by intraoperative navigation and execution. The preoperative planning software enables 3D virtual implant positioning based on the patient s CT reconstructed digital bone model and bony landmarks. The shoulder navigation application software then matches the patient s digital bone model and landmarks to the intraoperative landmarks registered by the surgeon, allowing an image-based navigation to follow. The surgeon uses the plan data as guidance to navigate and help position shoulder instruments and implants.

Reason for recall

Complaint identified issue with AI surgical planning software that may result in implant malalignment and/or decrease range of motion.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0145-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.