Philips Lumify Diagnostic Ultrasound System — Class I medical device recall Z-0152-2027
Ongoing recall by Philips Ultrasound, LLC, reported 2026-10-28, distributed nationwide. Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registr
| Recall number | Z-0152-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Philips Ultrasound, LLC, Bothell, WA, United States |
| Recall initiated | 2026-09-24 |
| Classified | 2026-10-16 |
| Reported by FDA | 2026-10-28 |
| Distributed | Nationwide (US) |
| Quantity | 13,206 |
| UPC / GTIN / UDI-DI | 00884838122116, 00884838124141 |
Product
Philips Lumify Diagnostic Ultrasound System
Reason for recall
Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/phone, there may be an unstable connection/may disconnect disrupting function. May cause diagnosis/treatment delay.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0152-2027. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-11-03. Not affiliated with or endorsed by FDA. Not medical or legal advice.