PERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture — Class II medical device recall Z-0153-2027
Ongoing recall by Ethicon, LLC, reported 2026-10-21, distributed nationwide. Silk sutures, indicated for use in cardiovascular, ophthalmic and neurological procedures, packaged with alcoh
| Recall number | Z-0153-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ethicon, LLC, San Lorenzo, PR, United States |
| Recall initiated | 2026-09-08 |
| Classified | 2026-10-14 |
| Reported by FDA | 2026-10-21 |
| Distributed | Nationwide (US) |
| Quantity | 1,764 |
| Reason classes | potency |
| UPC / GTIN / UDI-DI | 10705031047365, 30705031047369 |
| Expiration dates | 2026-09-30 |
Product
PERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture, 1X18" TF-5, Product Code: N271H
Reason for recall
Silk sutures, indicated for use in cardiovascular, ophthalmic and neurological procedures, packaged with alcohol-based fluid, not part of product design, that if discovered prior to use may cause surgical delay. If used, fluid could cause reduced suture tensile strength or knot security, tissue irritation/inflammatory tissue response, and treatment failure, including wound dehiscence/bleeding.