Data Foundry

Medical device recalls 2025

IV Administration sets — Class I medical device recall Z-0154-2026

Ongoing recall by ICU Medical, Inc., reported 2025-10-29, distributed nationwide. IV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop

Recall numberZ-0154-2026
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmICU Medical, Inc., San Clemente, CA, United States
Recall initiated2025-10-02
Classified2025-10-23
Reported by FDA2025-10-29
DistributedNationwide (US)
Countries outside the USAU, CA, NZ
Quantity23,203
UPC / GTIN / UDI-DI10840619005068, 10840619045859, 10887709007823, 10887709015033, 10887709015040
Lot numbers14065820, 14110075, 14126963, 14129823, 14211300, B33359/10887709007823/14070544, B9213/10840619045859/14130197, B9732/10887709015033/14149636, B9733/10887709015040/14176779
Model numbers011-C7014, B33359, B9213, B9732, B9733

Product

IV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set (w/Shut Off), MicroClave, Clamp, REF: B9213; 70" (178 cm) 150 ml Burette Set (Clave, Shut-Off) w/3-Way Stopcock, Clave, Graduated Adapter, Rotating Luer, 1 Ext, REF: B9732; 80" (203 cm) Y-Type 150 ml, Burette Set (Clave, Shut-Off), w/3-Way Stopcock, Clave, Graduated Adapter, Rotating, Luer, 1 Ext, REF: B9733; 53 cm (21") Add-On 150 ml Burette Set (Clave, Shut-Off), Vented Cap, REF: 011-C7014

Reason for recall

IV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or air may be infused into the body.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0154-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.