Data Foundry

Medical device recalls 2026

RAPICIDE PA Part A High-Level Disinfectant — Class II medical device recall Z-0155-2027

Ongoing recall by Steris Corporation, reported 2026-10-21, distributed nationwide. If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal press

Recall numberZ-0155-2027
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmSteris Corporation, Mentor, OH, United States
Recall initiated2026-09-21
Classified2026-10-14
Reported by FDA2026-10-21
DistributedNationwide (US)
Quantity11,806 units
UPC / GTIN / UDI-DI00724995245955
Model numbersML020116

Product

RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116

Reason for recall

If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0155-2027. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-27. Not affiliated with or endorsed by FDA. Not medical or legal advice.