DxU 840m Iris — Class II medical device recall Z-0160-2027

Ongoing recall by Beckman Coulter, Inc., reported 2026-10-21, distributed nationwide. The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replace

Recall numberZ-0160-2027
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBeckman Coulter, Inc., Miami, FL, United States
Recall initiated2026-08-12
Classified2026-10-15
Reported by FDA2026-10-21
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BH, BR, CA, CH, CN, CO, CW, CZ, DE, EC, ES, FR, GB, GH, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JO, JP, KR, KW, LB, LT, MA, MX, MY, NL, NZ, OM, PA, PH, PL, PT, QA, RO, RS, SA, SG, SK, TH, TW, UA, VN, ZA
Quantity2,088
Reason classeschemical contamination, packaging
UPC / GTIN / UDI-DI15099590748647, 15099590748654
Serial numbers840M
Model numbersC49513, C76947

Product

DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513

Reason for recall

The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.