IQ200 SPRINT — Class II medical device recall Z-0161-2027
Ongoing recall by Beckman Coulter, Inc., reported 2026-10-21, distributed nationwide. The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replace
| Recall number | Z-0161-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Beckman Coulter, Inc., Miami, FL, United States |
| Recall initiated | 2026-08-12 |
| Classified | 2026-10-15 |
| Reported by FDA | 2026-10-21 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AR, AT, AU, BD, BE, BH, BR, CA, CH, CN, CO, CW, CZ, DE, EC, ES, FR, GB, GH, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JO, JP, KR, KW, LB, LT, MA, MX, MY, NL, NZ, OM, PA, PH, PL, PT, QA, RO, RS, SA, SG, SK, TH, TW, UA, VN, ZA |
| Reason classes | chemical contamination, packaging |
| UPC / GTIN / UDI-DI | 10837461001256, 10837461001300, 10837461001355 |
| Serial numbers | IQ200 |
| Model numbers | 700-3322, 700-3322RFB, 700-3325, 700-3326, 700-3340, 700-3345, 700-3347, 700-3370, 700-3375, 700-3375RFB, 700-3376, B48997, B48998, B48999, C10683, C10684 |
Product
IQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325, 700-3326, B48997; IQ200 SELECT, REF: 700-3340, 700-3345, 700-3347, B48998, C10684; IQ200 ELITE, REF: 700-3370, 700-3375, 700-3375RFB, 700-3376, B48999, C10683
Reason for recall
The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.