Aviator Drill Bit — Class II medical device recall Z-0163-2013
Terminated recall by Stryker Spine, reported 2012-11-07, distributed nationwide. Aviator drill bits are 10 mm longer than the Reflex Hybrid drill bits and are not interchangeable which could
| Recall number | Z-0163-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Stryker Spine, Allendale, NJ, United States |
| Recall initiated | 2012-07-05 |
| Classified | 2012-10-29 |
| Reported by FDA | 2012-11-07 |
| Terminated | 2013-01-28 |
| Distributed | Nationwide (US) |
| Quantity | 6,154 units |
| Model numbers | 48770610, 48770612, 48770616 |
Product
Aviator Drill Bit (10 mm, 12 mm, 14 mm, 16 mm, Manufactured by: Stryker Spine SAS Z.1 Marticot -33610- Cestas France +33 (0) 5.57.97.06.30 http://www.stryker.com Distributed in the USA by Stryker Spine 2 Pearl Ct., Allendale NJ 07401-1677 USA +1-201-750-8000 Drill bits are reusable instruments, sterilized prior to use, used in conjunction with an Aviator Drill Guide to provide a pathway for screw insertion. The aviator Anterior Cervical Plating (ACP) System is intended for use as an aid in cervical spinal fusion and is intended for unilateral fixation.
Reason for recall
Aviator drill bits are 10 mm longer than the Reflex Hybrid drill bits and are not interchangeable which could lead to over-drilling by as much as 10 mm.