Data Foundry

Medical device recalls 2012

Aviator Drill Bit — Class II medical device recall Z-0163-2013

Terminated recall by Stryker Spine, reported 2012-11-07, distributed nationwide. Aviator drill bits are 10 mm longer than the Reflex Hybrid drill bits and are not interchangeable which could

Recall numberZ-0163-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmStryker Spine, Allendale, NJ, United States
Recall initiated2012-07-05
Classified2012-10-29
Reported by FDA2012-11-07
Terminated2013-01-28
DistributedNationwide (US)
Quantity6,154 units
Model numbers48770610, 48770612, 48770616

Product

Aviator Drill Bit (10 mm, 12 mm, 14 mm, 16 mm, Manufactured by: Stryker Spine SAS Z.1 Marticot -33610- Cestas France +33 (0) 5.57.97.06.30 http://www.stryker.com Distributed in the USA by Stryker Spine 2 Pearl Ct., Allendale NJ 07401-1677 USA +1-201-750-8000 Drill bits are reusable instruments, sterilized prior to use, used in conjunction with an Aviator Drill Guide to provide a pathway for screw insertion. The aviator Anterior Cervical Plating (ACP) System is intended for use as an aid in cervical spinal fusion and is intended for unilateral fixation.

Reason for recall

Aviator drill bits are 10 mm longer than the Reflex Hybrid drill bits and are not interchangeable which could lead to over-drilling by as much as 10 mm.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0163-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.