Williams Cystoscopic Injection Needles — Class II medical device recall Z-0163-2027
Ongoing recall by Cook Incorporated, reported 2026-10-21, distributed in FL, IL, NM, TX, WA. The female Luer lock adapter component in the affected device lots may not have been correctly seated due to m
| Recall number | Z-0163-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Cook Incorporated, Bloomington, IN, United States |
| Recall initiated | 2026-09-16 |
| Classified | 2026-10-15 |
| Reported by FDA | 2026-10-21 |
| Distributed | FL, IL, NM, TX, WA |
| Countries outside the US | CA, CH, DE, ES, FR, IT |
| Quantity | 97 units |
| UPC / GTIN / UDI-DI | 00827002161122 |
| Lot numbers | 16613606, 16951548, 16951548X |
| Model numbers | 090001-S26 |
Product
Williams Cystoscopic Injection Needles, used for cystoscopic injection into the urethra, bladder neck, and bladder wall, Reference Part Number 090001-S26, Order Number (GPN) G16112
Reason for recall
The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of this issue include increased procedural time and inadequate treatment.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0163-2027. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-27. Not affiliated with or endorsed by FDA. Not medical or legal advice.