Data Foundry

Medical device recalls 2013

Aquatec Ocean VIP — Class II medical device recall Z-0167-2014

Terminated recall by Invacare Corporation, reported 2013-11-13. The snap-on fixtures of the backrest cushion on the Mobile Shower and Toilet Commode Aquatec Ocean VIP and Oce

Recall numberZ-0167-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmInvacare Corporation, Elyria, OH, United States
Recall initiated2013-08-31
Classified2013-11-05
Reported by FDA2013-11-13
Terminated2014-09-15
Countries outside the USCA
Serial numbers13A04000420, 13A04000421, 13A04005696, 13A04005697, 13A04005698, 13A04006199, 13A04006200, 14541797, 14541800, 14541803, 14541804, 14541805, 14541806, 14541807, 14541808, 14541809, 14541810, 14541812, 14541820, 14541821, 14541822, 14541823, 14541825, 14541836, 14541837 and 475 more
Model numbers1470793, 1471208, 1525707

Product

Aquatec Ocean VIP, Aquatec Ocean Dual-VIP, Aquatec Ocean E-VIP mobile shower and toilet commode with tilting function. Mobile shower and toilet commode with tilting function

Reason for recall

The snap-on fixtures of the backrest cushion on the Mobile Shower and Toilet Commode Aquatec Ocean VIP and Ocean Dual VIP do not withstand a required pull out force of 500 N, when the product is used in a reclined position and or tilted position. This may result in the snap-on fixture coming unlatched.