COBAS INTEGRA 800 Analyzer with software 9864 — Class II medical device recall Z-0168-2014
Terminated recall by Roche Diagnostics Operations, Inc., reported 2013-11-13, distributed nationwide. Roche COBAS INTEGRA 800 system, Software Version 9864.C2 does not perform necessary ISE Service Actions. Whene
| Recall number | Z-0168-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Roche Diagnostics Operations, Inc., Indianapolis, IN, United States |
| Recall initiated | 2013-09-16 |
| Classified | 2013-11-05 |
| Reported by FDA | 2013-11-13 |
| Terminated | 2015-05-12 |
| Distributed | Nationwide (US) |
| Quantity | 527 units |
| Reason classes | software |
| Model numbers | 04559126001, 04589459970, 28122474001, 28122474692 |
Product
COBAS INTEGRA 800 Analyzer with software 9864.C2 Cobas. COBAS Part Numbers 28122474001, 28122474692, 04559126001, 04589459970 corresponding to the Analyzer with or without closed tube sampling accessory A fully automated system for clinical chemistry analysis intended for the in vitro quantitative/qualitative determination of analytes in body fluids.
Reason for recall
Roche COBAS INTEGRA 800 system, Software Version 9864.C2 does not perform necessary ISE Service Actions. Whenever the ISE rack is removed, the ISE counters in the software are reset to zero and the corresponding Service Actions are not conducted automatically by the system. Additionally, certain maintenance activities that have to be performed manually by the operator are not requested on the sof