Data Foundry

Medical device recalls 2023

Breast Plastics Pack — Class II medical device recall Z-0173-2024

Ongoing recall by American Contract Systems, Inc., reported 2023-11-01, distributed nationwide. During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and A

Recall numberZ-0173-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmAmerican Contract Systems, Inc., Temple Terrace, FL, United States
Recall initiated2023-09-01
Classified2023-10-26
Reported by FDA2023-11-01
DistributedNationwide (US)
Quantity20 cases
UPC / GTIN / UDI-DI00191072166241
Lot numbers756221, 782221, 862221
Model numbersBBPL20D

Product

Breast Plastics Pack, REF BBPL20D

Reason for recall

During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0173-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.