Data Foundry

Medical device recalls 2025

Medical convenience kits Item number/ Product description AHLD15X… — Class II medical device recall Z-0173-2026

Ongoing recall by American Contract Systems Inc., reported 2025-10-22, distributed nationwide. Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the init

Recall numberZ-0173-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmAmerican Contract Systems Inc., Grimes, IA, United States
Recall initiated2025-09-04
Classified2025-10-16
Reported by FDA2025-10-22
DistributedNationwide (US)
Quantity32,433 total
Reason classeschemical contamination
UPC / GTIN / UDI-DI00191072224019, 00191072224767, 00191072238993, 00191072239051
Lot numbers7311311, 8188711, 8424611, 8439111
Serial numbers17555039, 17555044, 17555049, 17609166, 17609181, 17682598, 17742543, 17754395, 47140875, 47141919, 8499211, AKDV96B, EVBH28D, UPVD01K

Product

Medical convenience kits Item number/ Product description AHLD15X LABOR DELIVERY PACK - 213387 AKDV96B VAGINAL DELIVERY AKDV96B VAGINAL DELIVERY EVBH28D BIRTHING ROOM PACK UPVD01K VAGINAL DELIVERY PACK

Reason for recall

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.