Data Foundry

Medical device recalls 2025

Medical convenience kits Item Number/Description BUCY78E CYSTO PACK… — Class II medical device recall Z-0174-2026

Ongoing recall by American Contract Systems Inc., reported 2025-10-22, distributed nationwide. Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the init

Recall numberZ-0174-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmAmerican Contract Systems Inc., Grimes, IA, United States
Recall initiated2025-09-04
Classified2025-10-16
Reported by FDA2025-10-22
DistributedNationwide (US)
Quantity32,433 total
Reason classeschemical contamination
UPC / GTIN / UDI-DI00191072141415, 00191072159922, 00191072205537, 00191072213990, 00191072228215, 00191072228864, 00191072235350
Lot numbers7990511, 8016811, 8026111, 8043211, 8217311, 8346211, 8392411, 8444211, 8531411
Serial numbers17435184, 17483121, 17483122, 17483143, 17483144, 17485283, 17499185, 17598793, 17622370, 17681124, 17692649, 17705421, CMCP10X, IHCY10Y, MHCY35AE, PLBY52E, UDCY31Z, UICY28L

Product

Medical convenience kits Item Number/Description BUCY78E CYSTO PACK CMCP10X CYSTO PACK - 205972 IHCY10Y CYSTO PACK MHCY35AE CYSTO PACK- 212487 PLBY52E CYSTO PACK UDCY31Z CYSTO PACK - 206047 UICY28L MOR CYSTO PACK

Reason for recall

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.