Data Foundry

Medical device recalls 2020

Bien-Air Surgery REF 1600407-001 SN **** FOOTCTRL OSSEODOC — Class II medical device recall Z-0176-2021

Ongoing recall by Bien-Air Surgery Sa, reported 2020-10-28, distributed nationwide. The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor.

Recall numberZ-0176-2021
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBien-Air Surgery Sa, Le Noirmont, N/A, Switzerland
Recall initiated2020-09-11
Classified2020-10-21
Reported by FDA2020-10-28
DistributedNationwide (US)
Quantity1,120 foot
Serial numbers16C0017, 16F0013, 16I0009, 16K0017, 17F0039, 17I0015, 17L0013
Model numbers1600407, 1600407-001

Product

Bien-Air Surgery REF 1600407-001 SN **** FOOTCTRL OSSEODOC - Product Usage: intended to convert electrical energy to mechanically drive handpieces and microsaws used to cut and shape bones.

Reason for recall

The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0176-2021. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.