Medical convenience kits Item Number/Description UICL44P ADULT… — Class II medical device recall Z-0177-2026

Ongoing recall by American Contract Systems Inc., reported 2025-10-22, distributed nationwide. Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the init

Recall numberZ-0177-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmAmerican Contract Systems Inc., Grimes, IA, United States
Recall initiated2025-09-04
Classified2025-10-16
Reported by FDA2025-10-22
DistributedNationwide (US)
Quantity32,433 total
Reason classeschemical contamination
UPC / GTIN / UDI-DI00191072194268, 00191072234018
Lot numbers8003911, 8317711
Serial numbers68244081, UICL44R

Product

Medical convenience kits Item Number/Description UICL44P ADULT CENTRAL LINE INSERTION SUPPLY KIT UICL44R ADULT CENTRAL LINE INSERTION SUPPLY KIT

Reason for recall

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.