Allen Life Assist Beach Chair — Class II medical device recall Z-0179-2023

Terminated recall by Baxter Healthcare Corporation, reported 2022-11-09, distributed in TX, WA, WI. the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a res

Recall numberZ-0179-2023
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBaxter Healthcare Corporation, Deerfield, IL, United States
Recall initiated2022-09-20
Classified2022-11-03
Reported by FDA2022-11-09
Terminated2026-08-04
DistributedTX, WA, WI
Countries outside the USDE, JP, KR, SE, ZA
Quantity1 units
UPC / GTIN / UDI-DI00615521004125
Lot numbers22-123
Model numbers104002, A-91501

Product

Allen Life Assist Beach Chair, BCSSP-BEACH CHAIR , REF A-91501, #104002. For use with orthopaedic procedures using the Fowlers or beach chair position.

Reason for recall

the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.