Data Foundry

Medical device recalls 2020

NanoClave Connector — Class II medical device recall Z-0182-2021

Terminated recall by ICU Medical, Inc., reported 2020-10-28, distributed nationwide. Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lo

Recall numberZ-0182-2021
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmICU Medical, Inc., San Clemente, CA, United States
Recall initiated2020-07-06
Classified2020-10-22
Reported by FDA2020-10-28
Terminated2023-05-11
DistributedNationwide (US)
Countries outside the USBE, BR, CA, DE, FR, IE, JP, MX, PL, SA
Reason classespackaging
UPC / GTIN / UDI-DI10887709051017
Lot numbers4447754, 4549051

Product

NanoClave Connector; 1 unit per pouch, 100 multivac pouches per case;UDI (01)10887709051017(17)241201(30)100(10)4549051; (01)10887709051017(17)241101(30)100(10)4447754 CLAVE Neutron is a normally closed, bidirectional connector intended for use as an accessory to an Intravascular catheter placed in the vein or artery.

Reason for recall

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0182-2021. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.