Data Foundry

Medical device recalls 2013

Applied Medical's Inzii¿ 12/15mm Retrieval System — Class II medical device recall Z-0183-2014

Terminated recall by Applied Medical Resources Corp, reported 2013-11-20, distributed nationwide. Applied Medical is conducting a voluntary recall of the Inzii¿ 12/15mm retrieval system. During shipment, the

Recall numberZ-0183-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmApplied Medical Resources Corp, Rancho Santa Margarita, CA, United States
Recall initiated2013-10-21
Classified2013-11-08
Reported by FDA2013-11-20
Terminated2014-09-15
DistributedNationwide (US)
Countries outside the USAE, AT, AU, BE, BR, CA, CH, CL, CO, CZ, DE, DK, EC, ES, FI, FR, GB, HK, IE, IL, IN, IT, JP, LB, MX, MY, NL, NO, NZ, PE, PL, PT, SA, SE, SG, ZA
Quantity29,506 units
Reason classessterility
Model numbersCD004

Product

Applied Medical's Inzii¿ 12/15mm Retrieval System. Applieds Inzii Tissue Retrieval Bag, Model Number CD004, is classified as a Class II device. It was cleared to market via 510(k) filing K060051. The retrieval bag is a disposable specimen retrieval bag intended for use as receptacle for the collection of tissue, organs, and calculi during laparoscopic procedures. Applieds CD004 has a shelf life of three years. It is provided sterile and packaged in a heat sealed Tyvek pouch.

Reason for recall

Applied Medical is conducting a voluntary recall of the Inzii¿ 12/15mm retrieval system. During shipment, the retrieval system packaging has the potential to become punctured with small holes, which could compromise the sterile barrier. The likelihood of this situation to occur and result in patient harm is highly unlikely; however, out of an abundance of caution for patient safety and a commitmen