Artis zeego systems — Class II medical device recall Z-0183-2015
Terminated recall by Siemens Medical Solutions USA, Inc, reported 2014-11-19, distributed nationwide. A cable inside the C-arm systems may have been routed in an inappropriate manner. Improper routing of the cabl
| Recall number | Z-0183-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Siemens Medical Solutions USA, Inc, Malvern, PA, United States |
| Recall initiated | 2014-09-23 |
| Classified | 2014-11-07 |
| Reported by FDA | 2014-11-19 |
| Terminated | 2017-05-08 |
| Distributed | Nationwide (US) |
| Quantity | 164 |
| Reason classes | device malfunction |
| Serial numbers | 1000-1599 |
| Model numbers | 10280959 |
Product
Artis zeego systems, model number 10280959, with serial numbers 1000-1599 : Artis zeego is a dedicated angiography system developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures.
Reason for recall
A cable inside the C-arm systems may have been routed in an inappropriate manner. Improper routing of the cable may result in increased wear over time. Without additional measures, the potential exists for a cable inside the C-arm to break which could result in restricted functionality or system failure and an ongoing procedure could be terminated. Additionally, a subgroup of affected C-arm systems will receive a replacement of the cable harness.