FERNO POWER X1 AMBULANCE COT- Item #0015807/UDI-DI 00190790002183… — Class II medical device recall Z-0184-2023
Ongoing recall by Ferno-Washington Inc, reported 2022-11-16, distributed nationwide. Batteries could become unstable when not properly maintained and/or subjected to repeated drops, power washing
| Recall number | Z-0184-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ferno-Washington Inc, Wilmington, OH, United States |
| Recall initiated | 2022-10-07 |
| Classified | 2022-11-04 |
| Reported by FDA | 2022-11-16 |
| Distributed | Nationwide (US) |
| Countries outside the US | BR, CL, CN, PE, PL |
| Quantity | 1,922 units |
| UPC / GTIN / UDI-DI | 00190790002183, 00190790004064, 00190790005474 |
| Serial numbers | 18F003862, 18F003863, 18F003864, 18F003865, 18F003866, 18F003867, 18F003868, 18F003869, 18F003870, 18F003871, 18F003872, 18F003873, 18F003874, 18F003875, 18F003876, 18F003877, 18F003878, 18F003879, 18F003880, 18F003881, 18F003882, 18F003883, 18F003884, 18F003893, 18F003894 and 475 more |
| Model numbers | 0015807, 0731403, 0822484/0822483 |
Product
FERNO POWER X1 AMBULANCE COT- Item #0015807/UDI-DI 00190790002183 with 36V Li-Ion Batteries Model Number #0822484/0822483 Note: Transcend Stair Chair Powertraxx (Item #0731403/UDI-DI 00190790004064) not included in this action).
Reason for recall
Batteries could become unstable when not properly maintained and/or subjected to repeated drops, power washing directly into battery contacts or abuse, may result in a fire and/or injury
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0184-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.