Optima NM/CT 640 — Class II medical device recall Z-0188-2015
Terminated recall by GE Healthcare, LLC, reported 2014-11-19, distributed nationwide. GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the
| Recall number | Z-0188-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | GE Healthcare, LLC, Waukesha, WI, United States |
| Recall initiated | 2014-10-14 |
| Classified | 2014-11-07 |
| Reported by FDA | 2014-11-19 |
| Terminated | 2015-02-27 |
| Distributed | Nationwide (US) |
| Countries outside the US | AR, AT, AU, BE, BG, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, HK, HR, HU, ID, IE, IL, IN, IR, IT, JP, KR, KW, LB, LT, LV, MA, MT, MX, MY, NL, NO, NZ, PA, PE, PH, PK, PL, PT, QA, RO, RU, SA, SE, SG, SK, TH, TN, TR, TW, UA, UY, VE, ZA |
| Quantity | 29 |
| Model numbers | H3100YA, H3100YC |
Product
Optima NM/CT 640, Models H3100YA, H3100YC. Product Usage: The GE Optima NM/CT 640 system is intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine (NM) system using a variety of scanning modes supported by various acquisition types, and a CT component which is intended specifically for enabling attenuation correction and anatomical localization on SPECT studies
Reason for recall
GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems.