Data Foundry

Medical device recalls 2014

Optima NM/CT 640 — Class II medical device recall Z-0188-2015

Terminated recall by GE Healthcare, LLC, reported 2014-11-19, distributed nationwide. GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the

Recall numberZ-0188-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGE Healthcare, LLC, Waukesha, WI, United States
Recall initiated2014-10-14
Classified2014-11-07
Reported by FDA2014-11-19
Terminated2015-02-27
DistributedNationwide (US)
Countries outside the USAR, AT, AU, BE, BG, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, HK, HR, HU, ID, IE, IL, IN, IR, IT, JP, KR, KW, LB, LT, LV, MA, MT, MX, MY, NL, NO, NZ, PA, PE, PH, PK, PL, PT, QA, RO, RU, SA, SE, SG, SK, TH, TN, TR, TW, UA, UY, VE, ZA
Quantity29
Model numbersH3100YA, H3100YC

Product

Optima NM/CT 640, Models H3100YA, H3100YC. Product Usage: The GE Optima NM/CT 640 system is intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine (NM) system using a variety of scanning modes supported by various acquisition types, and a CT component which is intended specifically for enabling attenuation correction and anatomical localization on SPECT studies

Reason for recall

GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0188-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.