Data Foundry

Medical device recalls 2013

SIGMA SPECTRUM Infusion Pump with Master Drug Library component — Class II medical device recall Z-0189-2014

Terminated recall by Baxter Healthcare Corp., reported 2013-11-20, distributed nationwide. Baxter Healthcare Corporation is recalling the LCD display screen on the SIGMA SPECTRUM Infusion Pump with Mas

Recall numberZ-0189-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBaxter Healthcare Corp., Deerfield, IL, United States
Recall initiated2013-09-24
Classified2013-11-08
Reported by FDA2013-11-20
Terminated2016-11-09
DistributedNationwide (US)
Countries outside the USCA
Quantity346 units
Serial numbers1002325, 1018432, 1018436, 1018737, 433713, 435563, 704056, 707290, 715315, 724216, 740127, 752318, 752447, 753675, 754947, 759766, 765934, 770034, 772017, 773470, 776507, 781016, 781969, 783352, 783454 and 66 more

Product

SIGMA SPECTRUM Infusion Pump with Master Drug Library component (SHARP LCD screen, 55074-2 revision B) Product Usage: Intended to be used as a visual aid during the controlled administration of intravenous fluids. These fluids may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy.

Reason for recall

Baxter Healthcare Corporation is recalling the LCD display screen on the SIGMA SPECTRUM Infusion Pump with Master Drug Library for not meeting standards for withstanding an electrostatic discharge.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0189-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.