Atellica CH 930 Analyzer — Class II medical device recall Z-0194-2019
Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2018-10-24, distributed nationwide. Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1.
| Recall number | Z-0194-2019 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Siemens Healthcare Diagnostics, Inc., Tarrytown, NY, United States |
| Recall initiated | 2018-06-29 |
| Classified | 2018-10-18 |
| Reported by FDA | 2018-10-24 |
| Terminated | 2021-08-12 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AT, AU, BE, BR, CA, CH, CZ, DE, DK, ES, FR, GB, GR, HR, IT, JP, KR, MX, MY, NL, NO, PL, SE, SG |
| Quantity | 115 |
| Reason classes | potency, software |
Product
Atellica CH 930 Analyzer (SMN 11067000) . The Atellica" Solution is a multi-component system for in vitro diagnostic testing of clinical specimens.
Reason for recall
Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks¿ and Ancillary Packs when more than one (1) lot of specific Reagent kits are loaded on the analyzer at the same time. 2. Sample barcodes may be incorrectly read 3. The Atellica IM 1300 Analyzer and Atellica IM 1600 Analyzer may not run CEA dilutions at 50x and 100x. The Analyzer may post a Sample Integrity Error when 50x or 100x dilutions are ordered for Carcinoembryonic Antigen (CEA). Dilution factors lower than 50x and neat samples are not affected by this behavior. 4. The Atellica CH 930 Analyzer may continue to display the default reporting units after the reporting units have been changed. 5. The Atellica CH 930 Analyzer may not display results for sample if certain assays are configured to auto-dilute results above or below the assay measuring interval.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0194-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.