ROTOR ASSY-SW40 — Class II medical device recall Z-0201-2027
Ongoing recall by Beckman Coulter, Inc., reported 2026-10-28, distributed nationwide. Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into t
| Recall number | Z-0201-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Beckman Coulter, Inc., Indianapolis, IN, United States |
| Recall initiated | 2026-09-21 |
| Classified | 2026-10-21 |
| Reported by FDA | 2026-10-28 |
| Distributed | Nationwide (US) |
| Quantity | 92 units |
| Reason classes | device malfunction |
| UPC / GTIN / UDI-DI | 15099590415754, 15099590415822, 15099590415846, 15099590415860, 15099590416324 |
| Serial numbers | 24U15260, 26U15764 |
| Model numbers | A94468, A94469, A94471, A94516, A95765, A99833, A99836, A99839, A99842, A99846 |
Product
ROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)
Reason for recall
Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparation, rotor breakage may delay patient test results. Additionally, sample destruction may expose laboratory personnel to biohazardous materials.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0201-2027. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-11-03. Not affiliated with or endorsed by FDA. Not medical or legal advice.