ROTOR ASSY-SW41 — Class II medical device recall Z-0202-2027
Ongoing recall by Beckman Coulter, Inc., reported 2026-10-28, distributed nationwide. Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two
| Recall number | Z-0202-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Beckman Coulter, Inc., Indianapolis, IN, United States |
| Recall initiated | 2026-09-21 |
| Classified | 2026-10-21 |
| Reported by FDA | 2026-10-28 |
| Distributed | Nationwide (US) |
| Quantity | 336 units |
| Reason classes | device malfunction |
| UPC / GTIN / UDI-DI | 15099590415754, 15099590415822, 15099590415846, 15099590415860, 15099590416324 |
| Serial numbers | 24U15250, 26U15766 |
| Model numbers | A94468, A94469, A94471, A94516, A95765, A99833, A99836, A99839, A99842, A99846 |
Product
ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)
Reason for recall
Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.