CareLink Clinic — Class II medical device recall Z-0203-2027
Ongoing recall by Medtronic MiniMed, Inc., reported 2026-10-28, distributed nationwide. CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which t
| Recall number | Z-0203-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic MiniMed, Inc., Northridge, CA, United States |
| Recall initiated | 2026-09-11 |
| Classified | 2026-10-21 |
| Reported by FDA | 2026-10-28 |
| Distributed | Nationwide (US) |
| Quantity | 1 |
| Reason classes | software |
| Model numbers | MMT-7350 |
Product
CareLink Clinic, REF: MMT-7350
Reason for recall
CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ratio, Insulin Sensitivity Factor (ISF), and Max Recommended value for Rapid-acting insulin. Instead, the calculator continues to use the system-recommended values, resulting in incorrect rapid-acting insulin parameters. If an HCP or patient inputs these incorrect setting values into the patient's MiniMed Go app, the patient could deliver higher or lower insulin doses compared to their insulin needs, potentially leading to over/under delivery of insulin with a severity ranging from Catastrophic (death from severe hypoglycemia) to Minor (self-treatable hypoglycemia/hyperglycemia).
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0203-2027. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-11-03. Not affiliated with or endorsed by FDA. Not medical or legal advice.