Techno-Path Multichem AE Unassayed Level 2 REF: AE622X — Class II medical device recall Z-0206-2027
Ongoing recall by TECHNO-PATH MANUFACTURING LTD., reported 2026-10-28, distributed nationwide. A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific
| Recall number | Z-0206-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | TECHNO-PATH MANUFACTURING LTD., Ballina Tipperary, Ireland |
| Recall initiated | 2026-08-14 |
| Classified | 2026-10-22 |
| Reported by FDA | 2026-10-28 |
| Distributed | Nationwide (US) |
| Quantity | 123 units |
| Reason classes | specification failure |
| UPC / GTIN / UDI-DI | 05391523442817 |
| Lot numbers | 65705262 |
| Model numbers | AE622X |
Product
Techno-Path Multichem AE Unassayed Level 2 REF: AE622X
Reason for recall
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0206-2027. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-11-03. Not affiliated with or endorsed by FDA. Not medical or legal advice.