Data Foundry

Medical device recalls 2015

e — Class II medical device recall Z-0208-2016

Terminated recall by Siemens Medical Solutions USA, Inc., reported 2015-11-11, distributed nationwide. Potential failure mode related to the radial brake, where the detector was able to drift, led to a re-design o

Recall numberZ-0208-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc., Hoffman Estates, IL, United States
Recall initiated2015-06-26
Classified2015-11-04
Reported by FDA2015-11-11
Terminated2016-12-23
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BO, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DK, EC, EG, ES, FI, FR, GB, GR, HR, IE, IL, IN, IQ, IR, IT, JO, JP, KR, KW, LB, LK, LU, LV, MA, MX, MY, NL, NO, NP, NZ, OM, PE, PH, PK, PL, PT, QA, RS, SA, SE, SK, SV, TH, TN, TR, TW, UA, ZA
Quantity32,296 e.cam
Serial numbers10000, 100003, 10001, 10002, 10003, 10004, 10005, 10006, 10007, 10008, 10009, 1001, 10010, 10011, 10012, 10013, 10014, 10015, 10016, 10017, 10018, 10019, 1002, 10020, 10022 and 475 more

Product

e.cam Gamma Camera, an emission computed tomography system. It is intended to detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV.

Reason for recall

Potential failure mode related to the radial brake, where the detector was able to drift, led to a re-design of the brake redundancy circuit. The redesigned Brake Redundancy Circuit will force even a defective brake to engage in the unpowered state by applying a reverse charge to the brake and ensuring the brake has demagnetized.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0208-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.