Data Foundry

Medical device recalls 2026

Automated Impella Controller — Class I medical device recall Z-0208-2027

Ongoing recall by Abiomed, Inc., reported 2026-11-04, distributed nationwide. Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or so

Recall numberZ-0208-2027
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmAbiomed, Inc., Danvers, MA, United States
Recall initiated2026-09-18
Classified2026-10-28
Reported by FDA2026-11-04
DistributedNationwide (US)
Countries outside the USAE, BE, BR, CA, CH, CO, CZ, DE, ES, FI, FR, GB, HK, HR, IE, IN, IT, KR, KW, LU, MX, MY, NL, NO, PA, PK, PL, SA, SE, SG, SI
Quantity428 units
Reason classessoftware
UPC / GTIN / UDI-DI00813502010022, 00813502011272, 00813502011289, 00813502011296, 00813502011418
Serial numbersIC1001, IC1014, IC1029, IC1038, IC1042, IC1043, IC1061, IC1075, IC1096, IC1122, IC1179, IC1190, IC1202, IC1203, IC1224, IC1231, IC1238, IC1269, IC1280, IC1283, IC1284, IC1285, IC1289, IC1345, IC1346 and 403 more

Product

Automated Impella Controller, Packaged. Primary user control interface for the Impella Catheter, catheter-based heart pump.

Reason for recall

Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0208-2027. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-11-10. Not affiliated with or endorsed by FDA. Not medical or legal advice.