Impella Optical Controller — Class I medical device recall Z-0209-2027
Ongoing recall by Abiomed, Inc., reported 2026-11-04, distributed nationwide. Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or so
| Recall number | Z-0209-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Abiomed, Inc., Danvers, MA, United States |
| Recall initiated | 2026-09-18 |
| Classified | 2026-10-28 |
| Reported by FDA | 2026-11-04 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, BE, BR, CA, CH, CO, CZ, DE, ES, FI, FR, GB, HK, HR, IE, IN, IT, KR, KW, LU, MX, MY, NL, NO, PA, PK, PL, SA, SE, SG, SI |
| Quantity | 179 units |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 00813502010961, 00813502010978, 00813502010985, 00813502011425 |
| Serial numbers | IC10040, IC10043, IC10063, IC10064, IC10102, IC10103, IC10149, IC10150, IC10238, IC10239, IC10278, IC10302, IC10318, IC10341, IC10366, IC10367, IC10463, IC10589, IC10979, IC10983, IC10984, IC10985, IC11174, IC11299, IC11410 and 154 more |
Product
Impella Optical Controller, Packaged. Primary user control interface for the Impella Catheter, a catheter-based heart pump.
Reason for recall
Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.
Machine access