Philips Azurion Interventional Fluoroscopic X-Ray System — Class II medical device recall Z-0213-2022
Terminated recall by Philips North America Llc, reported 2021-11-17, distributed nationwide. When the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the
| Recall number | Z-0213-2022 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Philips North America Llc, Cambridge, MA, United States |
| Recall initiated | 2021-09-24 |
| Classified | 2021-11-09 |
| Reported by FDA | 2021-11-17 |
| Terminated | 2024-09-26 |
| Distributed | Nationwide (US) |
| UPC / GTIN / UDI-DI | 00884838054202, 00884838085251, 00884838085268, 00884838085275, 00884838085282, 00884838085350, 00884838085367, 00884838099203, 00884838099210, 00884838099227, 00884838099234, 00884838099241, 00884838099258, 00884838099265, 00884838099272 |
| Model numbers | 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228 |
Product
Philips Azurion Interventional Fluoroscopic X-Ray System, Software version 2.1.x, Model numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228. (Azurion 3 M12, Azurion 3 M15, Azurion 5 M12, Azurion 5 M20, Azurion 7 B12, Azurion 7 B20, Azurion 7 M12, Azurion 7 M20)
Reason for recall
When the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the same time, the geometry may stop reacting on movement requests (table lock-up) and X-ray imaging becomes unavailable.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0213-2022. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.