Data Foundry

Medical device recalls 2019

SORIN Heater-Cooler System 3T 120V~/60Hz REF 16-02-85 Description of… — Class II medical device recall Z-0220-2020

Terminated recall by LivaNova USA Inc., reported 2019-11-06, distributed nationwide. The firm has become aware that due to a chemical reaction, the disinfectant/water preservative used to disinfe

Recall numberZ-0220-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmLivaNova USA Inc., Arvada, CO, United States
Recall initiated2018-11-01
Classified2019-10-25
Reported by FDA2019-11-06
Terminated2022-08-04
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BG, BH, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, GY, HK, HR, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KR, KW, LB, LK, LT, LU, LV, MA, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UA, UY, VN, ZA
Quantity8,671
Reason classeschemical contamination
Serial numbers16S10027, 16S10743, 16S10772, 16S10958, 16S11455, 16S12004
Model numbers16-02-85

Product

SORIN Heater-Cooler System 3T 120V~/60Hz REF 16-02-85 Description of product: The St¿ckert Heater-Cooler System is an independent (i.e. independent of the water supply) 3-circuit-heating/cooling system. The three separate water circuits have been developed to serve as: " Two circuits for the patient supplied by one tank (for the heating/cooling blanket and the heat-exchanger associated with the oxygenator), common temperatures ranging from 2 degrees Celsius to 41 degrees Celsius, tank volume 6 liters. " One interchangeable heating/cooling circuit for cardioplegia heat-exchanger, temperatures ranging from 2 degrees Celsius to 10 degrees Celsius (cooling tank) and/or from 15 degrees Celsius to 41 degrees Celsius (heating tank). If required, the circuits for the patient and the circuit for cardioplegia can be switched off separately, in order to increase the activated functional group s heating and/or cooling performance. Product Usage: The Stockert Heater-Cooler System 3T is intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung machine featuring a separate temperature control for extracorporeal perfusion of durations of up to 6 hours.

Reason for recall

The firm has become aware that due to a chemical reaction, the disinfectant/water preservative used to disinfect the device to inhibit/limit the growth micro-organisms, may be rendered ineffective. In addition, the firm is releasing a design upgrade to reduce the risk of potential emission of aerosols from the 3T.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0220-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.