Data Foundry

Medical device recalls 2021

Transseptal Needle with Catheter — Class II medical device recall Z-0223-2022

Terminated recall by Cook Inc., reported 2021-11-24, distributed nationwide. Transseptal needles may contain rust on the interior and/or exterior of the needle, which may not be detected

Recall numberZ-0223-2022
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmCook Inc., Bloomington, IN, United States
Recall initiated2021-10-08
Classified2021-11-12
Reported by FDA2021-11-24
Terminated2024-02-14
DistributedNationwide (US)
Quantity4,789
UPC / GTIN / UDI-DI00827002192614

Product

Transseptal Needle with Catheter. RPN TSN-17-75.0-ENDRYS. GPN G19261. Coaxial set consisting of an outer catheter, curved-tip metal cannula, and a tapered-tip inner needle

Reason for recall

Transseptal needles may contain rust on the interior and/or exterior of the needle, which may not be detected by the user. Potential adverse events if an affected product is used ranges from increased procedural time to inflammatory reactions, which at their most severe include systemic reactions potentially leading to permanent impairment or death.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0223-2022. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.