BACT/ALERT VIRTUO system — Class II medical device recall Z-0228-2019
Terminated recall by bioMerieux, Inc., reported 2018-10-31, distributed nationwide. The BACT/ALERT VIRTUO Instrument Firmware version R2.0 allows relocation of the four (4) resident calibration
| Recall number | Z-0228-2019 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | bioMerieux, Inc., Durham, NC, United States |
| Recall initiated | 2018-06-27 |
| Classified | 2018-10-23 |
| Reported by FDA | 2018-10-31 |
| Terminated | 2020-07-21 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AT, AU, BE, BR, CA, CH, CL, CO, CR, DE, EE, ES, FR, GB, HK, HR, IL, IN, IT, JP, KR, KW, LB, MY, NL, PA, PH, PT, QA, RS, SA, SE, SG, SI, TH, VN, ZA |
| Quantity | 516 units |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 03573026369767 |
Product
BACT/ALERT VIRTUO system, A Unit, with Version R2.0 Firmware Product Usage: The BacT/ALERT VIRTU Microbial Detection System is an automated microbial test system capable of incubating, agitating, and continuously monitoring for the detection of aerobic, facultative, and anaerobic microorganism growth from blood and other normally sterile body fluids.
Reason for recall
The BACT/ALERT VIRTUO Instrument Firmware version R2.0 allows relocation of the four (4) resident calibration standards so that Field System Engineers (FSE) can service the associated cells P24 - P27, where the calibration standards typically reside. If the calibration standards are not returned to the designated locations, an anomaly may occur.