SIGNA Architect system Product Usage: The Signa Architect system is a… — Class II medical device recall Z-0236-2019

Terminated recall by GE Medical Systems, LLC, reported 2018-10-31, distributed nationwide. A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a

Recall numberZ-0236-2019
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGE Medical Systems, LLC, Waukesha, WI, United States
Recall initiated2018-10-03
Classified2018-10-23
Reported by FDA2018-10-31
Terminated2022-05-27
DistributedNationwide (US)
Countries outside the USFR, KR, SE
Quantity4 units
Serial numbersU30RF0251B, U30RF0258B, U30RF0268B, U30RF0275B
Model numbers5388807-100

Product

SIGNA Architect system Product Usage: The Signa Architect system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Reason for recall

A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0236-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.