Computed Tomography Systems Revolution CT — Class II medical device recall Z-0242-2022
Ongoing recall by GE Healthcare, LLC, reported 2021-12-01, distributed nationwide. The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam
| Recall number | Z-0242-2022 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | FDA mandated |
| Recalling firm | GE Healthcare, LLC, Waukesha, WI, United States |
| Recall initiated | 2021-06-30 |
| Classified | 2021-11-24 |
| Reported by FDA | 2021-12-01 |
| Distributed | Nationwide (US) |
| Quantity | 105 in |
Product
Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition
Reason for recall
The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the SmartStep display will show a value that is higher (up to 100 times) than the actual dose delivered.