XCELA PASV 5F DL 55CM IR-145 KIT PG — Class II medical device recall Z-0243-2018
Terminated recall by Angiodynamics Inc. (Navilyst Medical Inc.), reported 2017-12-20, distributed nationwide. Navilyst Medical, Inc. (NMI) the manufacturer of the Xcela PICC with PASV, BioFlo PICC with PASV and the BioFl
| Recall number | Z-0243-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Angiodynamics Inc. (Navilyst Medical Inc.), Glens Falls, NY, United States |
| Recall initiated | 2016-11-28 |
| Classified | 2017-12-13 |
| Reported by FDA | 2017-12-20 |
| Terminated | 2019-06-12 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, CA, ES, GB, NL |
| Quantity | 4,591 units |
| Reason classes | device malfunction |
| Lot numbers | 4899431 |
| Model numbers | 45-733 |
Product
XCELA PASV 5F DL 55CM IR-145 KIT PG, UPN H965457330, Catalog No. 45-733 The Xcela with PASV, BioFlo PICC with ENDEXO Technology, and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring.
Reason for recall
Navilyst Medical, Inc. (NMI) the manufacturer of the Xcela PICC with PASV, BioFlo PICC with PASV and the BioFlo PICC, conducted this recall to the end user level based on information received from Greatbatch Medical, the manufacturer of the ViaPeel PTFE Peelable Introducer. Greatbatch Medical determined that the products listed in their 11/11/2016 Recall Notification have the potential for the handles to detach from the sheath during use. NMI has confirmed that the affected sheaths, Greatbatch Model Number 10890-006, had been included in packaged Xcela and BioFlo PICC Kits.