Data Foundry

Medical device recalls 2026

Direxion Torqueable Microcatheter — Class II medical device recall Z-0244-2027

Ongoing recall by Boston Scientific Cork, Ltd., reported 2026-10-28, distributed nationwide. The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the oute

Recall numberZ-0244-2027
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBoston Scientific Cork, Ltd., Cork, Ireland
Recall initiated2026-09-16
Classified2026-10-22
Reported by FDA2026-10-28
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BE, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DK, EC, ES, FI, FR, GB, GR, HR, HU, IE, IL, IT, JO, JP, KW, LB, LT, MX, NL, NO, NZ, PA, PE, PL, PT, QA, RO, RS, SA, SE, SG, SI, SK, TN, ZA
Quantity357
UPC / GTIN / UDI-DI08714729839521
Lot numbers38224583, 38273637, 38649877, 38707975, 38745198, 40470668, 40470669, 40476288, 40476289, 40476304, 40476305, 40495080, 40504684, 40512595, 40516625, 40516626, 40523967, 40525596, 40531425

Product

Direxion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/155; Material Number: M001195450

Reason for recall

The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.