Ilizarov Wire Tensioner — Class II medical device recall Z-0245-2022
Ongoing recall by Medicalplastic S.R.L., reported 2021-11-24, distributed in TN. Medicalplastic s.r.l. is conducting a voluntary field action to remove a group of Ilizarov Wire Tensioners due
| Recall number | Z-0245-2022 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medicalplastic S.R.L., Milan, N/A, Italy |
| Recall initiated | 2021-06-17 |
| Classified | 2021-11-17 |
| Reported by FDA | 2021-11-24 |
| Distributed | TN |
| Quantity | 581 units |
| Reason classes | device malfunction |
| UPC / GTIN / UDI-DI | 08033201842072 |
| Lot numbers | 20HP01926, 20HP01927, 20HP01928, 20LP02812, 20LP02813, 20LP02815, 20LP02816, 20LP02818, 20LP02819, 20LP02827, 20MP03035, 20MP03037, 20MP03041, 20MP03048, 20MP03049, 20MP03077, 20MP03078, 20MP03123, 20MP03124, 20MP03125, 20MP03126, 21BP00265, 21BP00266, 21BP00271, 21BP00272, MP03036 |
| Serial numbers | D059, D065, D099, D114, D115, D119, D120, D128, D129, D146, D147, D149, D175, D198, D199, D258, D259, D277, D278, D282, D283, D287, D288, D293, D294 and 26 more |
| Model numbers | 71070341 |
Product
Ilizarov Wire Tensioner, Model 71070341, UDI (10)08033201842072(10)21HP02550(11)210831(21E434
Reason for recall
Medicalplastic s.r.l. is conducting a voluntary field action to remove a group of Ilizarov Wire Tensioners due to a manufacturing error. An internal washer of the Ilizarov Wire Tensioner has been made in a wrong plastic material that could wear and break during the sterilization process in autoclave. Consequently this could cause malfunctioning of the tensioner.