Data Foundry

Medical device recalls 2017

Bio-Stable 5F DL-55CM MST-70 KIT Non-Valved PG — Class II medical device recall Z-0248-2018

Terminated recall by Angiodynamics Inc. (Navilyst Medical Inc.), reported 2017-12-20, distributed nationwide. Navilyst Medical, Inc. (NMI) the manufacturer of the Xcela PICC with PASV, BioFlo PICC with PASV and the BioFl

Recall numberZ-0248-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAngiodynamics Inc. (Navilyst Medical Inc.), Glens Falls, NY, United States
Recall initiated2016-11-28
Classified2017-12-13
Reported by FDA2017-12-20
Terminated2019-06-12
DistributedNationwide (US)
Countries outside the USAU, CA, ES, GB, NL
Quantity4,591 units
Reason classesdevice malfunction
Lot numbers4908978
Model numbers45-883

Product

Bio-Stable 5F DL-55CM MST-70 KIT Non-Valved PG, UPN H965458830, Catalog No. 45-883 The Xcela with PASV, BioFlo PICC with ENDEXO Technology, and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring.

Reason for recall

Navilyst Medical, Inc. (NMI) the manufacturer of the Xcela PICC with PASV, BioFlo PICC with PASV and the BioFlo PICC, conducted this recall to the end user level based on information received from Greatbatch Medical, the manufacturer of the ViaPeel PTFE Peelable Introducer. Greatbatch Medical determined that the products listed in their 11/11/2016 Recall Notification have the potential for the handles to detach from the sheath during use. NMI has confirmed that the affected sheaths, Greatbatch Model Number 10890-006, had been included in packaged Xcela and BioFlo PICC Kits.