Sysmex XE-5000 Automated Hematology System — Class II medical device recall Z-0255-2013
Terminated recall by Sysmex America, Inc., reported 2012-11-14, distributed nationwide. There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4
| Recall number | Z-0255-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Sysmex America, Inc., Mundelein, IL, United States |
| Recall initiated | 2012-06-01 |
| Classified | 2012-11-06 |
| Reported by FDA | 2012-11-14 |
| Terminated | 2016-08-22 |
| Distributed | Nationwide (US) |
| Quantity | 763 units |
| Model numbers | 063-7591-7 |
Product
Sysmex XE-5000 Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe, Japan; Sysmex America, One Nelson C. White Parkway, Mundelein, IL 60060; catalog #063-7591-7 The Sysmex XE-5000 is an automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XE-5000 performs analyses using the following methods: RF/DC Detection Method, Sheath Row DC Detection Method, and Flow Cytometry Methods using a Semiconductor Laser. Particle characterization and identification is based on detection of forward scatter, fluorescence and adaptive cluster analysis. Using the same reagents as the XE-2100, the XE-5000 automatically classifies cells from blood and body fluids and carries out all processes automatically from aspiration of the sample to outputting the results.
Reason for recall
There is a variation of reticulocyte counts between XE Series (XE-2100, XE-5000) and XT-Series (XT-2000i, XT-4000i) hematology analyzers that may result in a retic counts that are an average of 21% lower on the XE-Series because of a low bias.