Data Foundry

Medical device recalls 2017

Azurion 3 M12 — Class II medical device recall Z-0255-2018

Terminated recall by Philips Electronics North America Corporation, reported 2017-12-27, distributed nationwide. When using the downscale option with the Allura R9 system or Azurion R1.1 system, the measurements performed u

Recall numberZ-0255-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips Electronics North America Corporation, Andover, MA, United States
Recall initiated2017-08-18
Classified2017-12-18
Reported by FDA2017-12-27
Terminated2019-09-26
DistributedNationwide (US)
Quantity157

Product

Azurion 3 M12, Azurion 3 M15, Allura R9 7 M12, Allura R9 M20, Azurion 7 M12, Azurion 7 M20 The Allura Xper series is intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures ( such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP) Non-vascular interventions such as drainages, biopsies and vertebroplastics procedures The Azurion series (within the limits of the used Operation Room table) are intended for use to perform: Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures. Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures.

Reason for recall

When using the downscale option with the Allura R9 system or Azurion R1.1 system, the measurements performed using the QA Basic Measurement tool will not be correctly exported to the external DICOM destination.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0255-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.