Data Foundry

Medical device recalls 2021

Quantum 2000 Electrosurgical Generators Cooper Surgical Part Number:… — Class II medical device recall Z-0261-2022

Ongoing recall by CooperSurgical, Inc., reported 2021-11-24, distributed nationwide. Coagulate function may not operate with the use of the Hand Switch, failure of the device to coagulate tissue

Recall numberZ-0261-2022
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmCooperSurgical, Inc., Trumbull, CT, United States
Recall initiated2021-09-29
Classified2021-11-18
Reported by FDA2021-11-24
DistributedNationwide (US)
Countries outside the USAE, BO, CA, CL, CN, CR, DE, DO, EC, GB, HK, ID, IN, IR, IT, JO, KE, LK, MX, MY, NG, NI, NZ, PH, QA, SG, TH, TR, TT, TW, VN, ZA
Quantity193 units
Serial numbers0112010, 0114011, 0209001, 0212004, 0214002, 0214007, 0309031, 0312021, 0313007BFIX, 0313010B, 0411002, 0413003B, 0609010FIX, 0612001B, 0710001, 0710003, 0711008, 0711008FIX, 0713011C, 0713018C, 0809009, 0811017, 0813032CFIX, 0813033C, 091402 and 34 more
Model numbers909075

Product

Quantum 2000 Electrosurgical Generators Cooper Surgical Part Number: 909075

Reason for recall

Coagulate function may not operate with the use of the Hand Switch, failure of the device to coagulate tissue may lead to a prolonged procedure requiring medical intervention to control any intraoperative bleeding and complete the case.

Machine access