Data Foundry

Medical device recalls 2015

Samsung GM60A-40S Intended for use in generating radiographic images… — Class II medical device recall Z-0263-2016

Terminated recall by NeuroLogica Corporation, reported 2015-11-18, distributed in MA, MO, NC, OH. When the system is driving over a non-conductive floor and then over a conductive, grounded part, proper contr

Recall numberZ-0263-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmNeuroLogica Corporation, Danvers, MA, United States
Recall initiated2015-09-14
Classified2015-11-10
Reported by FDA2015-11-18
Terminated2017-01-19
DistributedMA, MO, NC, OH
Quantity7 devices
Serial numbers50EAM9BG200001A, 50EAM9BG200002P, 50EAM9BG200003D, 50EAM9BG200004X, 50EAM9BG200005N, 50EAM9BG200006E, 50EAM9BG200007F

Product

Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.

Reason for recall

When the system is driving over a non-conductive floor and then over a conductive, grounded part, proper control of the drive behavior may not be possible. The system can be stopped by releasing the drive handle or by pressing the emergency switch. Proper behavior is re-enabled when the system is rebooted.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0263-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.