Data Foundry

Medical device recalls 2012

Quadra S Offset Broach Handle The offset broach handle is intended to… — Class II medical device recall Z-0267-2013

Terminated recall by Medacta Usa Inc, reported 2012-11-14, distributed in CO, NV, TX. Medacta USA Inc has initiated a recall on the Quadra S Offset Broach Handle because of the potential of the ri

Recall numberZ-0267-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMedacta Usa Inc, Camarillo, CA, United States
Recall initiated2008-10-13
Classified2012-11-08
Reported by FDA2012-11-14
Terminated2012-11-09
DistributedCO, NV, TX
Countries outside the USBE
Quantity12 units
Lot numbers085119, 085120, 085243, 085244
Model numbers122, 123, 124, 125

Product

Quadra S Offset Broach Handle The offset broach handle is intended to attach to the broaches of different sizes as the surgeon broaches the femoral bone in preparation for the implantation of the femoral stem.

Reason for recall

Medacta USA Inc has initiated a recall on the Quadra S Offset Broach Handle because of the potential of the rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to lock/unlock the broaches.