Biomet 3i Narrow Right Angle Square Driver Tip — Class II medical device recall Z-0267-2016
Terminated recall by Biomet 3i, LLC, reported 2015-11-18, distributed in AK, AZ, CA, CO, CT, DC, FL, GA…. RASQ3N, a square, abutment screw driver tip has been labeled and distributed as RASH3N, a hexed, abutment scre
| Recall number | Z-0267-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Biomet 3i, LLC, Palm Beach Gardens, FL, United States |
| Recall initiated | 2015-08-10 |
| Classified | 2015-11-12 |
| Reported by FDA | 2015-11-18 |
| Terminated | 2016-10-12 |
| Distributed | AK, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, KY, LA, MA, MD, MI, MS, NC, NE, NJ, NY, OH, OR, PA, SD, TN, TX, UT, VA, WA |
| Countries outside the US | AR, AU, BE, CA, CO, GR, HK, IE, IT, KR, KW |
| Quantity | 103 units |
| Reason classes | labeling |
| Lot numbers | 1184554 |
| Model numbers | RASQ3N |
Product
Biomet 3i Narrow Right Angle Square Driver Tip. Driver tips are attached to a surgical drilling unit to carry components such as implant mounts, healing abutments and screws for placement and removal. For dental and restorative procedures.
Reason for recall
RASQ3N, a square, abutment screw driver tip has been labeled and distributed as RASH3N, a hexed, abutment screw driver tip. Additionally, the RASQ3N received the RASH3N's laser marking "RASH3N" during manufacturing.