Hamilton-C6 — Class II medical device recall Z-0267-2023
Ongoing recall by Hamilton Medical AG, reported 2022-11-30, distributed nationwide. Due to a malfunction related to the backlight on the ventilator screen. The screen goes black for approximatel
| Recall number | Z-0267-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Hamilton Medical AG, Domat/Ems, N/A, Switzerland |
| Recall initiated | 2022-09-22 |
| Classified | 2022-11-18 |
| Reported by FDA | 2022-11-30 |
| Distributed | Nationwide (US) |
| Reason classes | device malfunction |
| UPC / GTIN / UDI-DI | 07630002808590 |
| Serial numbers | 14784, 4833, 5663, 5679, 5681, 5939, 6075, 6093, 6149, 6157, 6159, 6212, 6219, 6231, 6274, 6276, 6280, 6309, 6320, 6327, 6349, 6350, 6416, 6464, 6480 and 358 more |
| Model numbers | 160021 |
Product
Hamilton-C6
Reason for recall
Due to a malfunction related to the backlight on the ventilator screen. The screen goes black for approximately 2-3 seconds, but it brightens up again when the backlight switches on again.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0267-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.