Data Foundry

Medical device recalls 2012

AC Powered adjustable hospital bed — Class II medical device recall Z-0270-2013

Terminated recall by Primus Medical LLC, reported 2012-11-14, distributed nationwide. The "Retaining washers & Clevis Pins" provided with the AC -powered adjustable hospital beds may fail under ce

Recall numberZ-0270-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPrimus Medical LLC, Boardman, OH, United States
Recall initiated2011-02-02
Classified2012-11-08
Reported by FDA2012-11-14
Terminated2014-05-14
DistributedNationwide (US)
Quantity342 beds
Reason classesdevice malfunction
Serial numbers001, 002, 003, 004, 005, 006, 007, 010, 013, 014, 015, 016, 017, 018, 021, 022, 023, 024, 025, 026, 027, 028, 029, 030, 031 and 316 more
Model numbersTL500

Product

AC Powered adjustable hospital bed, Prime Bed TL500, Distributed by: Primus Medical LLC, 8401 Southern Blvd, Boardman, OH 44512. Usage: Hospital Bed.

Reason for recall

The "Retaining washers & Clevis Pins" provided with the AC -powered adjustable hospital beds may fail under certain circumstances. The pins at the top of the leg were disengaging from the channel. This may cause the bed frame to come off tracks or collapse.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0270-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.