AC Powered adjustable hospital bed — Class II medical device recall Z-0270-2013
Terminated recall by Primus Medical LLC, reported 2012-11-14, distributed nationwide. The "Retaining washers & Clevis Pins" provided with the AC -powered adjustable hospital beds may fail under ce
| Recall number | Z-0270-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Primus Medical LLC, Boardman, OH, United States |
| Recall initiated | 2011-02-02 |
| Classified | 2012-11-08 |
| Reported by FDA | 2012-11-14 |
| Terminated | 2014-05-14 |
| Distributed | Nationwide (US) |
| Quantity | 342 beds |
| Reason classes | device malfunction |
| Serial numbers | 001, 002, 003, 004, 005, 006, 007, 010, 013, 014, 015, 016, 017, 018, 021, 022, 023, 024, 025, 026, 027, 028, 029, 030, 031 and 316 more |
| Model numbers | TL500 |
Product
AC Powered adjustable hospital bed, Prime Bed TL500, Distributed by: Primus Medical LLC, 8401 Southern Blvd, Boardman, OH 44512. Usage: Hospital Bed.
Reason for recall
The "Retaining washers & Clevis Pins" provided with the AC -powered adjustable hospital beds may fail under certain circumstances. The pins at the top of the leg were disengaging from the channel. This may cause the bed frame to come off tracks or collapse.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0270-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.